TWS-N15 General Purpose Noise Source: some progress on the design, and some tests

Many design of amateur noise sources have been published, they all have there benefits and shortcomings, but it is mostly the lack of a calibrated test source that makes it difficult for the hobbyist to employ a cheap, home-made noise source for the ever-so-critical noise figure measurements of the amplifiers and mixers he may build.
Having recently acquired a good calibrated source, HP 346B, I have decided to make available a noise source design and build so many, and calibrate them, in order to make calibrated sources, up to, say 1.5 or 2 GHz, available for everyone at a really low price. Thinking about USD 45 per piece which should be fine for everyone dealing with such measurements.
Requirements are pretty simple, it should be a ‘low-frequency’ replacement of the HP 346B, about 15 dB ENR output, flatness, preferably, within 0.5 dB, starting from 10 MHz. It should also have high output return loss (low SWR), which can’t change significantly from on to off state. Last but not least, powered by the ubiquitous 28 VDC noise source drive signal common to most noise figure analyzers.

The bill of materials, combined with the desired target price, won’t allow the use of a custom-made noise diode. But this is not really a disadvantage for the intended frequency range and purpose.

Three main tasks will have to be solved, to get this thing (called, the TWS-N15 Noise Source) to work:

(1) First and foremost, we need a noise generator circuit; these require a bias current, typically 5-15 mA.

(2) We need a temperature-compensated current source; currently experimenting with discrete Zener and TL431 based circuits. Current should not vary significantly, if input voltage is changing from, say, 27.5 to 28.5 VDC.

(3) Mechanical package. This is quite important, because the noise source needs to be well shielded, and put in a sturdy case – these sources tend to float around labs, and are often dropped, or dragged down from the bench by the heavy cables attached. Output connector will be SMA, because I have a large supply of really nice quality SMA print connectors.

The draft schematic (current source not shown, cirucit works, but still need better characterization, and possibly, some improvement). Note that the output attenuator will set the ENR level. Might need to adjust this a bit; for the time being, anything from 10 to 15 dB ENR will be fine.

tws noise source schematic

This is the small cirucit, build on a 0.6 mm FR4 board. Traces were simply carved out with a knive…..

2 ghz noise source

Here a few plots from the VNA characterization. Not bad for a start. Input return loss is at least 15 dB. This will still be improved. Using 0603 SMD resistors for the critical section.
tws noise smith chart

ON and OFF traces are shown – virtually, no effect on the return loss.
tws noise rfl loss on off

tws noise swr

Some tests of the noise output, really much better than I thought. Bias current has some impact on flatness; low frequency end is determined by the first capacitor (between diode and attenuator). This will need some further tweaking to get it as flat as possible, but no rocket science.
To measure the noise output flatness and level quickly, there is no also an automated test rig here, which measures bias current and on-off noise figure differences.

noise source test1

noise source test1 log

Next steps… will need to decide if the 0.6 mm board is the best option (these boards allow rather narrow 50 Ohm traces; most of them, I machine using a router mill – no etched traces), or if a commercially made board of regular dimensions would be the better option. Cost ist not really a concern, because the noise source only needs little space. Stay tuned.

HP (Agilent Keysight) 346B Noise Source: finally, a calibrated ENR standard, and a temperature compensated current source

Today, a rather ordinary envelope arrived, still it feels a bit like xmas, because of the contents….
346b envelope

…. a HP (Agilent Keysight) 346B Noise source, with nominal 15 dB ENR. I have long been looking for one, at a reasonable price, and finally scored this unit on xbay.

346b

The calibration sticker shows good flatness, especially, in the 0.01-1 GHz region, which I need most, it is perfectly flat.

346b enr cal

Prior to having it re-calibrated, a good opportunity to look inside. There are two sections: the current source, and the RF noise source assembly (which is hermetically sealed, and you better don’t touch!).

346b parts

These are some close-ups of the 00346-60001 power supply and current regulator board.

346b top side

346b bottom side

The current regulator inside of the 346B has always been a big mystery to me, because no schematic has been published by HP, in any of the service manuals. How it works, check out the schematic. The incomming 28 V (which is provided by the noise figure meter) is converted to a square wave, about 7-8 kHz, using a LM311H comparator. This is then converted to a negative voltage, about -20 V, absolute value varies a bit with loading condition.
The negative voltage is then used to sink current from the noise source assembly. The current setting of my unit is about 18.7 mA, programmed by the “10 Ohm” resistor.
Why the negative voltage? To simplify the design of the noise diode.

00346-60001 346B noise source schematic

Why did HP use a 5V6 Zener for the current reference, well, this is fairly obvious, when looking at the datasheet of such diodes.

bzx85 zener data

Around 5.6 V, the temperature coefficient virtually vanishes (this is why such voltages are also used for voltage reference circuits). The other Zener diode, unfortunately, I was not able to identify. It has 24 V drop, fair enough, any regular Zener should to the job to keep the base bias of the output constant.

For a quick burn-in, the source has now been connected to a 8970A Noise Figure Meter, and output appears to be very steady, less then 0.1 dB drift over a few hours. More noise measurements and calibration tasks to come, let me know if you need any noise sources measured, might be able to help.